Meet Our Protein Characterization Problem Solver

MAURICE

Reproducible, quantitative analysis of identity, purity and heterogeneity profiles for your therapeutic proteins

Your workhorse station for purity, identity and heterogeneity analysis

Maurice innovates the conventional capillary electrophoresis technology to automate your protein profiling either by size or charge. Need cIEF and CE-SDS data for your mAbs, ADCs, vaccines or virus-like particles? Maurice does it all. Just pop in one of the preassembled cartridges, drop in your sample vials or a 96-well plate, and hit start. You will complete method development in a day; and get high-resolution, reproducible data in minutes.

 

What Maurice can do for you?


  • cIEF Charge Application
  • CE-SDS Size Application
  • Absorbance Detection
  • Fluorescence Detection
  • Onboard Mixing for Sample Prep
 

See How Maurice Makes Protein Analysis Easy

 

Applications

 

Specifications

Scroll more >>>

Description cIEF Specification CE-SDS Plus Specification Turbo CE-SDS
Typical Separation Time 6-10 minutes (molecule-dependent) Reduced: 25 minutes; Non-reduced: 35 minutes Reduced: 5.5 minutes; Non-reduced: 8 minutes
Detection Capability Pre-focusing: 1500 V, focusing: 3000 V Separation: 5750 V UV Abs at 220 nm
Sample Injections per Cartridge 100 guaranteed, 200 maximum 500 maximum for CE-SDS PLUS 100 guaranteed, maximum 25 batches
Maximum Sample Injections per Batch 100 48 96
pI/Sizing CV 1% ≤2% <2%
pI/Sizing Resolution 0.05 pl units (for wide range 3-10 ampholyte) ≥ 1.5 for NGHC/HC IgG Standard) ≥ 1.0 for NGHC/HC IgG Standard)
Sensitivity (LOD) 0.7 µg/mL (Native fluorescence)
3.0 µg/mL (Absorbance) (Values based on a monoclonal antibody)
0.3 µg/mL
(Value based on internal standard)
0.6 μg/mL (value based on internal standard)
Minimum Sample Volume 50 µL 100 µL
Sample Delivery Vacuum Electrokinetic
pI/Size Range 2.85-10.45 10-270 kDa
CV for Peaks >10% Composition ≤5% (Intra-batch), ≤6% (Inter-batch) N/A
Dynamic Range 2 logs
Linearity >0.995
Sample Tray Options 96-well plates or 48 vials
Power 100 V-240 V (AC), 50/60 Hz, 500 W
Temperature Control Range 4-25 °C
Dimensions 44cm H x 42cm W x 61cm D
Weight 46 kg (100 lb)
 

Publications

 

Screening of process-related impurities (top) and mAb stability post-degradation (bottom) to validate the quality of mAb.

 
 
 


Talk to Our Scientists / Request a Demo